7.4 Troubleshooting

746 VA Trace Analyzer / 747 VA Stand

7-41

•

Enter a new half-wave potential in the instrument and recalculate the results.

•

Electrolyte solution too old: make up a new one. Its working life with organic

additives may be as short as 1 day or less.

No peak found

•

The peak is only displaced: adjust the half-wave potential and recalculate.

•

The background current is too high: repolish the electrode.

•

The sample concentration is too low: increase sample volume.

•

The deposition time under

MEAS

in "OP.SEQUENCE" in the inverse voltam-

metry is too short: increase the time.

•

Electrolyte solution too old: make up a new one. Its working life with organic

additives may be as short as 1 day or less.

Peak is in the highest

µ

A range

•

The sample volume is too large: reduce the volume. Carry out the analysis

again.

•

The background current is too high: repolish the electrode.

•

The deposition time is too long: reduce the time.

•

Is the deposition potential correct?

Double peak

•

Check RDE and polish if necessary (see section 3.5).

•

If organic components are present: carry out a UV digestion or other suitable

sample preparation.

•

Electrolyte solution too old: make up a new one. Its working life with organic

additives may be as short as 1 day or less.

•

Try another measurement mode like AC1. If one substance is reversible and

the second one irreversible, only the reversible substance is detected by

AC1.

Standard addition peaks displaced

•

Standard solutions have been made too acidic.

•

Buffering capacity of the electrolyte is not sufficient: increase electrolyte vol-

ume.

•

Electrode surface overcharged: reduce sample volume.

No addition

•

Has the correct standard solution been used or is the concentration of the

solution too low: increase the volume of the standard addition or use a higher

concentration or reduce the sample amount accordingly.

•

If organic components are present: carry out a UV digestion or other suitable

sample preparation.

•

Concentration of the analyte too high: dilute.

•

Electrolyte solution too old: make up a new one. Its working life with organic

additives may be as short as 1 day or less.