5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-191
ADD)n
Page:
O, S
Stand:
M, A
Abbreviation:
AD
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. ADD)n .....
Function
End of the loop for standard addition.
Parameter
Range; Default value / Meaning
n
0...8; 0
Number of standard additions
This parameter defines the number of repetitions of the ADD loop, i.e.
the number of standard additions.
Description
The
ADD)n
instruction ends a loop opened with
(ADD
for standard
additions. This loop must include one of the dispensing instructions
ADD/M
,
ADD>M
or
PADD>M
for addition of the standard addition solution as
well as call-up of a sweep or segment. Any other additional instructions
can be incorporated.
Comments
see
(ADD
Example
see
(ADD
see also
(ADD
,
ADD/M
,
ADD>M
,
PADD>M