Metrohm 746 VA Trace Analyzer User Manual

Page 119

Advertising
background image

4 Introduction to the operation

746 VA Trace Analyzer / 747 VA Stand

4-34

4.3.4

Recording the calibration curve

Recording of the calibration curve involves the following steps:

1

Switch to dialog page "MONITORING"

Press the yellow <MONITORING> key. The dialog page "MONITORING"

whose header shows the loaded method

CALCRV

appears.

2

Switch the program run mode to

calibration

If the parameter

Run mode

is set to

determination

, proceed as follows:

Use the cursor keys to select the field

Run mode

.

Press the space bar to enter the edit mode.

Press the softkey

<calibr.>

and confirm this entry with the <ENTER>

key.

3

Entry of the sample volume
Enter the sample volume as follows:

Use the cursor keys to select the first field below the title

Sample

size/S0

.

Press the space bar to enter the edit mode.

Enter either

*

or

10 mL

(as preparation for performance of the determi-

nation in section 4.3.5) and confirm this entry with the <ENTER> key.

Although the sample volume for recording the calibration curve is not needed (the
volume of the calibration solutions is defined in the parameter
V0.CSoln), the
Sample size/S0 field must not be empty.

4

Start the determination

Press the red <START> key.

The status display in the header of the dialog page "MONITORING"

switches to

* BUSY *

, the start time of the measurement is entered un-

der

Determination

and the determination is given a standard name,

also formed from the start time. Further, during the measurement the

measurement window is superimposed which shows the current pro-

gram line being executed and the remaining line time.

Shortly after the start of the determination, the prompt to add 10 mL of

calibration solution

Pb-1

to the measuring vessel appears on the

screen:

5

Addition of 1

st

calibration solution

Remove stopper 25

25 and use a pipette to add 10 mL of calibration

solution

Pb-1

with a mass concentration of 2.5

µµg/L to the measuring

vessel through the pipetting opening. Press the <ENTER> key.

The screen shows the prompt to add 1 mL acetate buffer to the measur-

ing vessel:

Ъ

ЪДДД

ДДД MEAS.TASK MESSAGE ДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДД¿

¿

³

³

³

³

³

³

CSOLN Pb-1 10.000 mL > M

³

³

³

³

³

³

А

АДДД

ДДД <Enter>: continue ДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДЩ

Щ

Advertising