5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-242

HMDE

Page:

O, S

Stand:

M, A

Abbreviation:

H

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. HMDE ..... Drop size ........ Meas.cell ........

Function

Switch on Hanging Mercury Drop Electrode (HMDE).

Application

The HMDE is primarily used for very sensitive stripping voltammetry in
which the analyte species is not measured until it has first been
electrochemically enriched. Cyclic voltammetry and PSA are also
possible with the HMDE.

Parameter

Range; Default value / Meaning

Drop size

1...9; 4

Drop size

Setting of the size of the mercury drop.
The area of the drop is ca. 0.15 mm

2

(

1

) ... 0.60 mm

2

(

9

).

The set drop time

t.drop

important for certain parameter settings is

0 ms for the HMDE by definition.

Meas.cell

normal, EFTA; normal

Operation mode

Setting of the operation mode for the HMDE:

normal

Normal operation for MME in measuring vessel:
4 mercury drops of defined size are formed in succes-
sion at the MME. The last drop remains suspended

EFTA

Operation for EFTA cell (future application):
The valve for Hg outflow is opened until two drops have
been expelled. After a waiting time of 300 ms a new Hg
drop of the defined size

Drop size

is formed and re-

mains suspended.

Description

The HMDE instruction is used to form mercury drops of defined size at
the 6.1246.020 Multi-Mode Electrode. The last drop remains suspended
and thus forms the hanging mercury drop electrode at which a voltage
sweep can be performed.

t

F (Hg)

t.HMDE

Area (Hg)