Metrohm 746 VA Trace Analyzer User Manual

Page 509

Advertising
background image

7 Safety, Errors, Troubleshooting, Diagnosis, GLP

746 VA Trace Analyzer / 747 VA Stand

7-10

Error message

Meaning/Causes

Rectification

Exit

CURVES - ERROR 3:
No segment available

No segment data whatsoever

available

Start new measurement

E

CURVES - ERROR 4:
Selected segment not
found

The selected segment (VR) was

not found

Switch to available segment

E

CURVES - ERROR 5:
Can not execute,
missing data

Action can not be executed

owing to missing result data

Start new measurement

E

CURVES - ERROR 6:
Can not display
calibration curve

No calibration curve available

If really desired, start new measurement

with calibration

E

CURVES - ERROR 7:
Can not handle
meas.data with actual
method

The current method has been

so changed that the experi-

mental data curves can no

longer be displayed

Reload determination with original

method or start new measurement

E

CURVES - ERROR 8:
Unqualified peak
attribute for this
operation

The requested operation can

not be performed with the

selected peak (e.g.

<Modify

baseline>

with peak without

baseline)

Read remarks on selected peak on the

”RESULTS/Evaluation” page

E

CURVES - ERROR 9:
Can not find plot
parameters of segment

The segment curve can not be

displayed as no plot parame-

ters are available (e.g. as the

segment name has been

changed)

Reset segment name (the name may

not be changed after the determination

has been recorded)

E

DATA CARD - ERROR 1:
Hardware error

Hardware fault on data card or

unexpected write/read error

−

Check data card, reinsert and attempt

write/read operation again

−

If error reappears, repeat write/read

test with a different data card (if suc-

cessful, reformat old card)

−

If error reappears, inform Metrohm

service

E/C

DATA CARD - ERROR 2:
Card invalid or not
available

−

No data card inserted

−

Data card not inserted

properly

−

Battery of data card fully

discharged

−

Insert data card

−

Take out data card then reinsert and

push right in

−

Change battery of data card

E/C

DATA CARD - ERROR 3:
Battery word bad

Test pattern in RAM of data

card destroyed, i.e. data card

battery discharged

Change battery of data card

E/C

DATA CARD - ERROR 4:
Low battery charge
detected

Charge of data card battery low

−

Copy all data on data card to instru-

ment memory

−

Change battery

−

Recopy data to data card

E/C

DATA CARD - ERROR 6:
Invalid file descriptor

Data interchange with data

card not possible

Press Reset key

E/C

DATA CARD - ERROR 7:
Can not open more files
simultaneously

Data interchange with data

card not possible

Press Reset key

E/C

DATA CARD - ERROR 8:
File name already
exists

Error in changing file name:

Inputted new name is inadmis-

sible as a file with the same

name exists (overwrite protec-

tion)

Use a different file name or delete old

file

E/C

Advertising