297 stir, O, s, M, a – Metrohm 746 VA Trace Analyzer User Manual
Page 448

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-297
STIR
Page:
O, S
Stand:
M, A
Abbreviation:
ST
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. STIR ..... Rot.speed ......../min
Function
Switch on stirrer.
Parameter
Range; Default value / Meaning
Rot.speed
0, 200, 400, ..., 3000 /min; 2000 /min
Rotational speed
Setting of the number of revolutions of the stirrer per min.
Description
The
STIR
instruction is used to switch on the stirrer of the 747 VA Stand.
The stirrer comprises the 6.1246.010 Drive shaft and the 6.1204.090
Stirrer tip (see section 3.8).
The stirrer remains active until it is switched off by the
ØSTIR
instruction
or
END
in the main program on the "OP.SEQUENCE" page.
Comment
If the stirrer should be switched off automatically after a definite time, the
TSTIR
instruction can also be used.
see also