Metrohm 746 VA Trace Analyzer User Manual
Page 575
Advertising
8 Technical data
746 VA Trace Analyzer / 747 VA Stand
8-14
8.3.5
Handshake
The 746 VA Trace Analyzer offers the following possibilities to set the handshake
(see section 5.3.1):
No handshake (
none
)
The 746 VA Trace Analyzer neither checks handshake inputs (CTS, DSR, DCD) nor
sets handshake outputs (DTR, RTS).
Reduced hardware handshake (
HWshort
)
746 VA Trace Analyzer as receiver:
LF
RTS
DTR
RxD
RTS
DTR
RxD
746 VA Trace Analyzer as sender:
LF
RTS
CTS
TxD
RTS
CTS
TxD
DTR
DTR
The data flow can be interrupted by deactivating the CTS line.
Full hardware handshake (
HWfull
)
All handshake inputs are checked, all handshake outputs are set.
746 VA Trace Analyzer as receiver:
LF
RTS
DTR
RxD
RTS
DTR
RxD
Time
External device
746
Time
External device
Gerät
746
Time
External device
746
Advertising