Metrohm 746 VA Trace Analyzer User Manual
Page 218

5.4.5 COMMON VARIABLES
746 VA Trace Analyzer / 747 VA Stand
5-67
Softkeys
Meaning
Modify
column
Modify parameters column-by-column
Modify
line
Modify parameters line-by-line
Delete
line
Delete line
The line selected with the field cursor is deleted. The two fields
Resid.com.var
and
Remark
or
Workg.com.var
and
Remark
are
considered as lines here.
Load all
--->
Load common variables into working memory
The common variables permanently stored in the instrument are
loaded into the working memory where they can be used for formulae.
Save all
<---
Save common variables from the working memory
The common variables loaded in the working memory are perma-
nently stored in the instrument as
Resid.com.var
. This is particularly
advisable before loading an old determination as this area is then
overwritten with the common variables valid for this determination.
Delete
Resid.CV
Delete all instrument-resident common variables
Delete
Workg.CV
Delete all common variables in working memory
Parameter
Range; Default value / Meaning
Resid.com.var
-3.7e+33...-1.7e-33, 1.7e-33...3.7e+33;
Instrument-resident common variables C01...C18
Numeric variables common to different methods permanently stored
in the instrument.
Remark
16 characters;
Comment on common variable
Free alphanumeric entry.
Workg.com.var
-3.7e+33...-1.7e-33, 1.7e-33...3.7e+33;
Common variables C01...C18 in working memory
Numeric variables common to different methods stored in the working
memory.
Remark
16 characters;
Comment on common variable
Free alphanumeric entry.