7 Safety, Errors, Troubleshooting, Diagnosis, GLP

746 VA Trace Analyzer / 747 VA Stand

7-38

Standard addition curves are not reproducible or non-linear

•

Check program in the instrument (stirring time, etc.).

•

Check MME, change capillary if necessary (see section 3.4.8) or replace

sealing needle (see section 3.4.9).

•

The pipetting process was not correct: was the pipetting unit used properly?

•

Repeat the analysis again or try out automatic standard addition with Dosi-

mat, Dosino or 6.5611.010 Pipetting equipment.

•

Pipetting the standard solutions must be carried out by one and the same

person or with the same instrument or the same pipette.

•

Check and test pipetting process. When were the pipettes last calibrated

(GLP)?

•

Organic components interfere with the analysis: carry out a UV digestion or

other suitable sample preparation.

•

Are the calibration solutions too old?

•

Would a calibration curve be more suitable?

•

The linearity at the HMDE is naturally not as good as with the DME. The linear

range is in general no larger than 1 - 2 decades.

Peak displacement

•

Check and adjust the pH of the solution.

•

Check electrolyte composition and correct if necessary.

•

Organic components interfere with the analysis: carry out a UV digestion or

other suitable sample preparation.

•

Carry out a standard addition to check if the correct peak is present.

•

Enter a new half-wave potential in the instrument and recalculate the results.

•

Electrolyte solution too old: make up a new one. Its working life with organic

additives may be as short as 1 day or less.

No peak found

•

The peak is only displaced: adjust the half-wave potential and recalculate.

•

The sample concentration is too small: increase the sample volume.

•

Has the complex former been forgotten? (adsorptive voltammetry).

•

The deposition time under

Meas

in "OP.SEQUENCE" in the inverse voltam-

metry is too short: increase the time.

•

No Hg drops at the capillary: check MME. Adjust sealing needle (see section

3.4.4/3.4.5). If necessary, change capillary (see section 3.4.8) or replace

sealing needle (see section 3.4.9).

•

Electrolyte solution too old: make up a new one. Its working life with organic

additives may be as short as 1 day or less.

•

Are organic components present? Carry out a UV digestion or other suitable

sample preparation.