3 Installation

746 VA Trace Analyzer / 747 VA Stand

3-24

3.4.8

Changing the capillary

Contamination of the glass capillary can necessitate its replacement (see also sec-

tions 7.4.2 and 7.4.3). In such a case, proceed as follows:

1 Remove multi-mode electrode from 747 VA Stand

•

Unscrew FEP tubing 51

51 and 59

59 from the MME, disconnect electrode

cable 37

37 from MME.

•

Take multi-mode electrode 38

38 out of measuring head 41

41 while gently

tapping the MME to knock off any mercury drops on the glass capillary

into the measuring vessel.

•

Place multi-mode electrode 38

38 in a horizontal position in the electrode

holder 114

114 (see Fig. 15-3).

2 Unscrew slotted screw 96

96

•

Using a suitable coin, unscrew slotted screw 96

96 out of screw thread 98

98

until the contact surface of the black O-ring at the Plexiglas wall (thin,

black stripe) is just visible below the metal thread 98

98.

3 Dismantle old capillary

•

Position multi-mode electrode 38

38 vertically in the electrode holder 114

114

(see Fig. 15-1).

•

Undo retaining nut 108

108 completely by turning anticlockwise and lift up

until the lower part of the glass capillary 109

109 with the wide opening is

visible.

•

Gently tap the glass capillary 109

109 to knock off any residual mercury in the

wide opening into the MME.

•

Press the retaining nut 108

108 downward with one hand and with your other

hand take glass capillary 109

109 completely out of the mount.

4 Dispose of old capillary

•

Connect filling tubing 115

115 with the tubing coupling 118

118 to the two gas

wash bottles 117

117 and the vacuum pump (see Fig. 14).

•

Insert glass capillary 109

109 (capillary end) in the filling cone 116

116 of the filling

tubing 115

115.

•

Remove mercury from capillary with the vacuum pump.

5 Replenish mercury if necessary

Proceed as described in section 3.4.2.

6 Mount new capillary

Proceed as described in section 3.4.3.

7 Fill capillary

Proceed as described in section 3.4.4 or section 3.4.5.