Metrohm 746 VA Trace Analyzer User Manual

Page 515

Advertising
background image

7 Safety, Errors, Troubleshooting, Diagnosis, GLP

746 VA Trace Analyzer / 747 VA Stand

7-16

Error message

Meaning/Causes

Rectification

Exit

MALLOC - ERROR 1:
System heap, not enough
memory

Insufficient system storage

space available (typical case:

sending large or many files)

This error message is acknowledged

automatically after 10 s waiting time.

r/a

MALLOC - ERROR 2:
Determination heap, not
enough memory

Insufficient storage space

available for determinations

Press <a> (abort)

Create storage space by deleting old

determinations

Restart copying

r/a

MALLOC - ERROR 3:
Method heap, not enough
memory

Insufficient storage space

available for methods

Press <a> (abort)

Create storage space by deleting old

methods

Restart copying

r/a

MALLOC - ERROR 4:
Storage heap, not
enough memory

Insufficient storage space

available for methods or

determinations

Press <a> (abort)

Create storage space by deleting old

methods or determinations

Restart copying

r/a

MALLOC - ERROR 5:
Can not realloc, not
enough memory

Insufficient storage space

available

Press <a> (abort)

Create storage space by deleting old

methods or determinations

Restart copying

r/a

MALLOC - ERROR 6:
Last powerdown failed,
malloc data incon-
sistent

Inadmissible power failure

during data storage (this error

message appears when the

instrument is switched on)

Press <r> (retry)

Store data from method and determi-

nation memory on data card

Reinitialize method and data memory

r/a

MALLOC - ERROR 7:
System heap out of
memory

Insufficient system storage

space available

Wait a short time, then press <r>

(retry)

If error reappears, press <a> (abort);

an instrument reset is initiated auto-

matically

Note:

If this error message appears, the

<PRINT SCREEN> key must not be

pressed (danger of crash)

r/a

MEASURE - ERROR 1:
Missing method

The method called with

Method.call

is missing in the

method memory

Load method in method memory or

correct method name

E/S

MEASURE - ERROR 2:
Faulty method

Programming error on

”OP.SEQUENCE” or

”SEGMENTS” page

Check programs on the two pages

and correct

If the programming error is not found,

press the softkey

<Check method>

(method test) on the ”MONITORING”

page and note error messages in

error window

Correct programs in accordance with

the error messages

E/S

MEASURE - ERROR 3:
Missing sample

With automatic program call

Auto.samples = on

, no

sample available in sample

memory

With manual sample call

Auto.samples = off

, no

value for

S0

available

Enter new sample data or set

Auto.samples

to

off

Enter

Sample size/S0

on the

”MONITORING” page

E/S

MEASURE - ERROR 4:
Too many sweeps

The maximum of 60 sweep

blocks has been exceeded

Reduce number of sweeps

E/S