5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-260
Comments
•
The
PSAMODE
measurement mode can be used only with the
electrode types
RDE
and
HMDE
and must be employed together with
the
PSWEEP
instruction.
•
The electrode type
RDE
or
HMDE
must be entered before the
PSAMODE
measurement mode.
•
The measurement time
t.meas
for the measurement of the voltage
drop is divided into 11 windows of different size in which 64 meas-
ured values are recorded each time (resulting in a total of 704 meas-
ured values per segment). The first measurement window has a size
of 2.944 ms, each subsequent window is twice as large as the pre-
ceding window.
•
With mercury film electrodes based on the RDE, it is advisable to
deposit the substances with stirring on the electrode. As in the
PSWEEP
instruction, only an enrichment without stirring is possible (no
stirring is allowed during measurement), the
MEAS
instruction must be
inserted. The enrichment time with stirring is defined by its line time,
in general
U.meas = U.start
is selected for the measurement volt-
age.
Example
see
PSWEEP
see also
PSWEEP, MEAS, RDE, HMDE