Metrohm 746 VA Trace Analyzer User Manual

Page 539

Advertising
background image

7 Safety, Errors, Troubleshooting, Diagnosis, GLP

746 VA Trace Analyzer / 747 VA Stand

7-40

7.4.4

Problems with RDE

There are various different types of RDE:

GC: Glassy carbon electrode, can be polished with aluminium oxide, glossy

surface.

Au: Gold electrode, can be polished with aluminium oxide, mat surface.
UT: Ultra-trace graphite electrode, can be scraped and polished with alumin-

ium oxide, mat surface.

Ag: Silver electrode, can be polished with aluminium oxide, mat surface.
Pt:

Platinum electrode, looks very similar to Ag, mat surface.

Low background current or unstable base line

•

The electrode surface must be repolished.

•

Has the correct RDE been used?

•

Exchange RDE.

•

Check concentration of electrolyte and pH of the solution.

•

Has the electrode been conditioned (e.g. with

CONDC

) ?

•

Check initial and final potential of the analysis.

•

If the ion concentration in the solution is too high: dilute the electrolyte.

•

If the concentration to be determined is considerably higher than expected:

reduce sample volume.

•

The background current is normally higher if RDE is used in place of MME.

•

A background current of several 100 nA is entirely possible.

•

Electrolyte solution too old: make up a new one. Its working life with organic

additives may be as short as 1 day or less.

Standard addition curves are not reproducible

•

Check program in the instrument (stirring time, etc.).

•

Check RDE (see section 3.5).

•

The pipetting process was not correct: was the pipetting unit used properly?

•

Repeat the analysis again or try out automatic standard addition with Dosi-

mat, Dosino or 6.5611.010 Pipetting equipment.

•

Pipetting the standard solutions must be carried out by one and the same

person or with the same instrument or the same pipette.

•

Check and test pipetting process. When were the pipettes last calibrated

(GLP)?

•

Organic components interfere with the analysis: carry out a UV digestion or

other suitable sample preparation.

•

Are the calibration solutions too old?

•

Would a calibration curve be more suitable?

Peak displacement

•

Check and adjust pH of the solution.

•

Check electrolyte composition and correct if necessary.

•

Organic components interfere with the analysis: carry out a UV digestion or

other suitable sample preparation.

Advertising