7 program instructions, Description in the – Metrohm 746 VA Trace Analyzer User Manual
Page 339

5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-188
Description
In the
AC2MODE
measurement mode, a sinusoidal alternating voltage with
a small, constant amplitude and low frequency is superimposed on a
stepwise rising direct voltage ramp. The 2
nd
harmonic wave of the alter-
nating current component I produced by the alternating voltage is
measured as a function of the voltage U. The nature of the measurement
is illustrated below using the SMDE as an example.
Adjustable parameters:
AC2MODE
t.step
U.ampl
Modul.freq.
Ph.angle
Prep.cycles
Meas.cycles
SWEEP
U.start
U.end
U.step
t.drop
t
U
t
U.step
U.start
Prep.
cycles
Meas.
cycles
t.step
Ph.angle = 0
°
F (Hg)
t.modul = 1/Modul.freq.
U.ampl
× √
2
Ph.angle > 0
°
Ph.angle
Ph.angle < 0
°
90
º
180
º
270
º
360
º
Ph.angle
-180
º
-90
º
90
º
180
º
270
º
360
º
(I
n+
)
(current addition)
(I
n–
)
(current subtraction)
t.meas
(I
n+
)
t.meas
(I
n–
)
t.meas
(I
n–
)
t.meas
(I
n+
)
Area (Hg)