Metrohm 746 VA Trace Analyzer User Manual

Page 526

Advertising
background image

7.3 Error messages

746 VA Trace Analyzer / 747 VA Stand

7-27

Error message

Meaning/Causes

Rectification

Exit

SUBSTANCES - ERROR 3:
Param. Mass conc.
missing

Missing value for the

Mass

conc.

parameter

Enter

Mass conc.

parameter on the

”SUBSTANCES” page

Esc

SUBSTANCES - ERROR 4:
Param. Scope f./r.half
only with linear

The

f.half/r.half

pa-

rameter is admissible only with

a linear baseline

Change

Scope

or

Type

parameter on

the ”SUBSTANCES” page

Esc

SUBSTANCES - ERROR 5:
Param. Scope
f./r.double missing
counterpart

Missing counterpart for the

evaluation with

f.double

or

r.double

; in double peak

evaluation the first peak must

be evaluated with

f.double

and the second peak with

r.double

Check

Scope

parameter for both

substances and change

Esc

SUBSTANCES - ERROR 6:
Param. Scope
f./r.double direction
conflict

Wrong assignment of the two

peaks in the double peak

evaluation; the first peak (in

sweep direction) must be

evaluated with

f.double

, the

second with

r.double

Check

Scope

parameter for both

substances and change

Esc

SUBSTANCES - ERROR 7:
Undefined error

Non-specified error on the

”SUBSTANCES” page

Check the ”SUBSTANCES” page

Esc

VA STAND - ERROR 1:
Missing device

Connection to 747 VA Stand

broken:

−

Connecting cable not

plugged in

−

Wrong address

−

Hardware fault at VA Stand

Check connection to 747 VA Stand:

−

Plug in connecting cable

−

Correct address and restart 746 VA

Trace Analyzer

−

Run diagnostic routine for VA Stand

E/S

VA STAND - ERROR 2:
Working electrode
faulty

Fault at MME working electr.:

−

No connection

−

Too less mercury

−

Bad mercury quality

−

Dirty or clogged capillary

−

Dirty or damaged sealing

needle

−

No inert gas pressure

−

Dirty electrode body

−

Fault at reference electrode

−

No solution in measuring

vessel

−

No electrolyte in measuring

vessel

−

Tapping strength in DME

operation too weak

Fault at RDE working electrode:

−

No connection

−

Electrode tip not firmly

screwed on to the drive shaft

−

Contaminated electrode

surface

Check MME:

−

Connect reference electrode with WE

cable

−

Fill MME with mercury (see section

3.4.2 and 3.4.7)

−

Clean MME (see section 3.4.9) and

use ultrapure mercury (w

≥

0.99999)

or filtered mercury

−

Change capillary (see section 3.4.8)

−

Clean MME (see section 3.4.9) and

replace sealing needle if need be

−

Check inert gas connection (see

section 3.3.3)

−

Rub the MME body dry, clean with

distilled water of dilute acid and dry

with a lint-free cloth

−

see

VA STAND - ERROR 4

−

Add measuring solution

−

Add a few drops of KCl 3 mol/L to the

measuring solution

−

Turn slotted screw 71

71 on the 747 VA

Stand during operation in an anti-

clockwise direction until a drop falls

each time the tapper is triggered

Check RDE:

−

Connect reference electrode with WE

cable

−

Install RDE properly (see section 3.5)

−

Regenerate RDE (see section 3.5.2)

r/a