5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-299

SWEEP

Page:

S

Stand:

M, A

Abbreviation:

SW

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. SWEEP XXXXX U.start ........mV U.step ........mV

U.end ........mV Sweep rate XXXXXXXXmV/s

Function

Voltage sweep.

Parameter

Range; Default value / Meaning

t/s

display only

Time for voltage sweep

In the line time (which can not be edited), the time for the voltage
sweep is displayed and is calculated as follows:

(U.end - U.start)

Sweep time = –––––––––––––––––––––

Sweep rate

U.start

-3000...3000 mV; -

Initial voltage

Initial voltage for the voltage sweep.

U.end

-3000...3000 mV; -

Final voltage

Final voltage for the voltage sweep. This voltage is retained after the
end of the sweep until a new voltage is applied or until it is switched off
by

ØMEAS

or

END

in the main program on the "OP.SEQUENCE" page.

U.step

2, 4, 6, 8, 10, 12 mV; 6 mV

Voltage step

Voltage step for direct current ramp.

Sweep rate

display only

Ramp slope

Here, the slope of the direct current ramp is displayed in mV/s. It is
calculated as follows:

U.step

Sweep rate = ––––––––

t.step

The parameter

t.step

originates from the measurement mode, which

must be entered before the

SWEEP

instruction.