Metrohm 746 VA Trace Analyzer User Manual

Page 337

Advertising
background image

5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-186

The phase dependent differences in the current measurements give
peak-shaped curves which can be evaluated using the baselines of type

linear

,

polynom

or

expon.

(see section 5.6.3).

I

U

U.peak

I.peak

Comments

•

The

AC1MODE

measurement mode can be used with all types of

electrodes together with the instructions

MEAS

,

SWEEP

,

DSWEEP

,

FSWEEP

or

TSWEEP

.

•

The electrode type

DME

,

SMDE

,

HMDE

or

RDE

must be entered before

the

AC1MODE

measurement mode.

•

With the

AC1MODE

measurement mode, the potentiostat always

operates in the operating mode "fast".

SWEEP

and

FSWEEP

are thus

identical for this mode.

•

The quantity

t.drop

important for the definition of

t.step

depends

on the type of electrode selected:

DME

t.drop = 20 ms

HMDE and RDE

t.drop = 0 ms

SMDE

t.drop = Drop.size

×× 40 ms

see also

AC2MODE, SWEEP, DME, SMDE, HMDE, RDE