Metrohm 746 VA Trace Analyzer User Manual

Page 382

Advertising
background image

5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-231

Adjustable parameters:

DPMODE

t.step
t.meas
t.pulse
U.ampl

SWEEP

U.start
U.end
U.step

F (Hg)

t.drop

t

U

t

U.step

U.start

I

U

t.meas

(I

1

)

U.peak

I.peak

U.ampl

t.step

t.pulse

t.meas

(I

2

) t.meas

(I

1

) t.meas

(I

2

)

I = I

2

- I

1

Comments

•

The

DPMODE

measurement mode can be used with all types of

electrodes together with the instructions

MEAS

,

SWEEP

,

DSWEEP

,

FSWEEP

or

TSWEEP

.

•

The electrode type

DME

,

SMDE

,

HMDE

or

RDE

must be entered before

the

DPMODE

measurement mode.

•

If the

DPMODE

measurement mode is used with the voltage sweep

SWEEP

and the electrode type

HMDE

, the potentiostat normally oper-

ates in the operating mode "slow". A switch is made to the operating
mode "fast" when the pulse time

t.pulse

< 30.05 ms.

•

If the

DPMODE

measurement mode is used with

SWEEP

and the

electrode type

SMDE

or

DME

, the setting of the potentiostat depends

on the measurement profile. Normally, a switch is also made from
"slow" to "fast" when the pulse time

t.pulse

< 30.05 ms. With very

small signals, however, the operating mode "slow" is used and with
very large signals the operating mode "fast", irrespective of the pulse
time.

•

If the

DPMODE

measurement mode is used with the voltage sweep

SWEEP

and the electrode type

RDE

, the potentiostat always operates in

the operating mode "fast".

•

If the

DPMODE

measurement mode is used with the voltage sweep

FSWEEP

, the potentiostat always operates in the operating mode "fast".

This mode is recommended especially for adsorptive stripping tech-
niques (e.g. Ni and Co with DMG).

•

The quantity

t.drop

important for the definition of

t.step

depends

on the type of electrode selected:

DME

t.drop = 20 ms

HMDE and RDE

t.drop = 0 ms

SMDE

t.drop = Drop.size

×× 40 ms

see also

SWEEP, FSWEEP, DME, HMDE, SMDE, RDE

Area (Hg)