7.4 Troubleshooting

746 VA Trace Analyzer / 747 VA Stand

7-35

7.4.2

Problems with DME/SMDE

Low background current or unstable baseline

•

The electrode drops irregularly: check MME.

Adjust sealing needle (see section 3.4.4/3.4.5). If necessary, change capillary

(see section 3.4.8) or replace sealing needle (see section 3.4.9).

•

Check electrolyte concentration and pH of the solution.

•

Check initial potential and final potential of the analysis.

•

If the ion concentration in the solution is too high: dilute the electrolyte.

•

If the concentration to be determined is considerably lower than anticipated:

increase sample volume or change mode (e.g. HMDE).

•

Has the sample been degassed? Degassing with nitrogen for at least 5 min is

recommended, for alkaline solutions approx. 10 min is recommended.

•

Check tapping mechanism on VA Stand. If tapping strength is too weak, turn

slotted screw 71

71 on the valve block during operation in an anticclockwise di-

rection until a drop falls each time the tapper is triggered.

•

Is the gas pressure correctly set (1 bar)?

•

Is the reference electrode sufficiently full (inside and outside, see section

3.6)?

Curves look like a "starry night"

•

The electrode drops irregularly: check MME.

Adjust sealing needle (see section 3.4.4/3.4.5). If necessary, change capillary

(see section 3.4.8) or replace sealing needle (see section 3.4.9).

•

Check contact between needle and capillary. If necessary, clean the MME

(see section 3.4.9).

•

Have stirring or degassing been switched off during the measurement?

•

If the electrode drops much too quickly: reduce

t.step

in program.

•

Check tapping mechanism on stand. If tapping strength is too weak, turn

slotted screw 71

71 on the valve block during operation in an anticclockwise di-

rection until a drop falls each time the tapper is triggered.

Standard addition curves are not reproducible

•

Check program in the instrument (stirring time, etc.).

•

Check MME, change capillary if necessary (see section 3.4.8).

•

The pipetting process was not correct: was the pipetting unit used properly?

•

Repeat the analysis again or try out automatic standard addition with Dosi-

mat, Dosino or 6.5611.010 Pipetting equipment.

•

Pipetting the standard solutions must be carried out by one and the same

person or with the same instrument or the same pipette.

•

Check and test pipetting process. When were the pipettes last calibrated

(GLP)?

•

Organic components interfere with the analysis: carry out a UV digestion or

other suitable sample preparation.

•

Are the calibration solutions too old?

•

Would a calibration curve or the internal standard method be more suitable?