5.5.3 CURVES

746 VA Trace Analyzer / 747 VA Stand

5-103

Substance-VR-Curve: XXXXXXXX-XX-smoothed

U.peak : XXXXXX mV dU.front: XXXXXX mV dU.rear: XXXXXX mV

I.peak : XXXXXX xA S.front: XXXXXX S.rear: XXXXXX

P.peak : XXXXXX xW

U.width: XXXXXX mV

Modify

column

Start

zoom

New

display

Reset

param.

Save

param.

Softkeys

Meaning

Modify

column

Modify display/plot parameters column-by-column

Modification of the parameters for the axis scaling. For details of the
display/plot parameters, see section 5.6.3.

Modifications made here initiate a recalculation and curve display on
closure of the entry. These changes are temporary, they are canceled
with

<Reset param.>

or after exit with <ESC>.

The changes are permanently incorporated in the method only if they
are stored with

<Save param.>

before exiting.

Start

zoom

Select zoom range

This softkey starts the selection of the zoom region (curve section
which should be enlarged).

First, the cursor keys <è>, <ç>, <é> and <ê> must be used
to define the top left corner point of the zoom region (marked by a
cross) as follows:

Each time these keys are pressed, the corner point moves in the
selected direction by 5 pixels. A finer shift of 1 pixel can be
achieved when the <SHIFT> key is pressed at the same time. In
both cases, the cursor keys are repetitive, i.e. the longer the cur-
sor key is pressed, the further the corner point is shifted.

Uppr/lft

corner

Once the top left corner point is in the desired position, it must be
fixed by pressing the softkey

<Uppr/lft corner>

or the <ENTER>

key. Next, the bottom right corner point of the zoom region is also
defined using the cursor keys. This automatically opens a frame
encompassing the region to be enlarged.

9

Display / Plot

––––––––––––––––––––––––––

I.scale ........

I.begin ........A

I.end ........A

U.div ........V/cm

U.begin ........V

U.end ........V

8