5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-296
Description
The
SSWEEP
instruction starts the voltage sweep defined in the parame-
ters.
U
t
U.start
U.end
U.step
t.step
Comments
•
The electrode type must be defined before the
SSWEEP
instruction.
Admissible are
DME
,
SMDE
,
HMDE
or
RDE
.
•
The measurement mode must be defined before the
SSWEEP
instruction. Admissible is only
DCTMODE
.
•
Apart from the
SSWEEP
instruction, no other SWEEP instruction is
allowed in the same segment.
•
If the final voltage
U.end
should not be switched off at the end of the
program, enter a suitable standby voltage
U.standby
in a
ØMEAS
instruction.
•
With the
SSWEEP
instruction maximum 1000 measured values can be
recorded.
see also
SWEEP, DSWEEP, FSWEEP, DME, SMDE, HMDE, RDE, DCTMODE