5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-189

The phase dependent differences in the current measurements give
curves which can be evaluated using the baselines of type

AC2 rel.

or

AC2 abs.

(see section 5.6.3).

U / 2

I.abs

I

U

Comments

•

The

AC2MODE

measurement mode can be used with all types of

electrodes together with the instructions

MEAS

,

SWEEP

,

DSWEEP

,

FSWEEP

or

TSWEEP

.

•

The electrode type

DME

,

SMDE

,

HMDE

or

RDE

must be entered before

the

AC2MODE

measurement mode.

•

With the

AC2MODE

measurement mode, the potentiostat always

operates in the operating mode "fast".

SWEEP

and

FSWEEP

are thus

identical for this mode.

•

The quantity

t.drop

important for the definition of

t.step

depends

on the type of electrode selected:

DME

t.drop = 20 ms

HMDE and RDE

t.drop = 0 ms

SMDE

t.drop = Drop.size

×× 40 ms

see also

AC1MODE, SWEEP, DME, SMDE, HMDE, RDE

Σ

(I

n+

– I

n–

)

n

I

=

I.rel

= I

f

+ I

r

I

f

I

r

n = Meas.cycles