5.6 Method Specifications

746 VA Trace Analyzer / 747 VA Stand

5-132

Parameter

Range; Default value / Meaning

U.width max

(U.width min)...400 mV; 200 mV

Maximum peak width

Maximum peak width for the assignment of a peak or a wave to a
substance. The peak width

aU.width

estimated in a first approxima-

tion must be smaller than

U.width max

for the substance to be

recognized (see adjacent diagram and section 6.6).

I.threshold

100...5000 pA; 200 pA

Limiting current value

Limiting current value for the assignment of a peak or wave to a
substance (for all measurement modes with the exception of
PSAMODE). With the aid of the 1

st

derivative of the peak curve (or the

2

nd

derivative of the wave curve), the difference

minmax

is multiplied

with an empirical factor f

MM

dependent on the measurement mode to

determine the estimated peak height

approx.peak

. This estimated

value must be larger than

I.threshold

for the substance to be

recognized (see also section 6.6).

With non-gaussian shaped peaks, the peak height approx.

peak estimated from the derivative can differ greatly from

the actual peak height

I.peak

. It may be necessary to

modify

I.threshold

for the results to be displayed or

suppressed in the desired manner.

aU.width
U.width min

U.width max

U.verify ± U.tol

aU.peak

approx.peak =

aU.max

aU.min

I.threshold

I

dI/dU

U

U

The substance recognition is
based on the recognition
parameters specified for the
individual substances. The
following conditions must be
met for a detected peak to
be assigned to a substance:

1.

aU.peak =
U.verify

±± U.tol

2.

aU.width > U.width min
aU.width < U.width max

3.

approx.peak >
I.threshold

minmax

minmax

×

f

MM

estimated

gaussian

peak