Backgnd) – Metrohm 746 VA Trace Analyzer User Manual
Page 349

5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-198
Example
A measurement with background compensation and standard addition
is normally programmed on the OP.SEQUENCE page with the following
basic structure:
:
DOS>M or DOS/M
Addition of supporting electrolyte
:
to measuring vessel
:
(BACKGND
Start of block for background compensation
SEGMENT Seg1
Call-up of the segment (BACKGND curve)
BACKGND)
End of block for background compensation
:
SMPL>M or SMPL/M
Sample addition to the measuring vessel
(ADD
Loop start for standard additions
(REP
Loop start for replications
SEGMENT Seg1
Call-up of the segment (measured curve)
REP)n
Loop end, n = number of replications
ADD>M, ADD/M or PADD>M
Constant or proportional standard addition
ADD)n
Loop end, n = number of standard additions
:
see also
BACKGND)