237 f>a, O, s – Metrohm 746 VA Trace Analyzer User Manual

Page 388

Advertising
background image

5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-237

F>A

Page:

O, S

Stand:

A

Abbreviation:

F

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. F>A ..... addition FEED -> AUX V.add ........mL

Function

Transfer solution from sample vessel (FEED) to the auxiliary vessel
(MEAS) on the 695 Autosampler.

Parameter

Range; Default value / Meaning

V.add

0.010...50.000 mL; 1.000 mL

Transfer volume

Solution volume that should be transferred from the sample vessel on
the sample rack of the 695 Autosampler to the auxiliary vessel.

Description

The

F>A

instruction transfers the solution volume specified under

V.add

from the sample vessel (FEED) to the auxiliary vessel (AUX) on the 695
Autosampler.

Comments

•

The

F>A

instruction can be used only in operation with the 695

Autosampler.

•

The auxiliary vessel (AUX) must be located on the sample rack
immediately after the sample vessel (FEED).

•

If the solution volume in the sample vessel <

V.add

, an error

message appears.

see also

A>M

,

F>M