237 f>a, O, s – Metrohm 746 VA Trace Analyzer User Manual
Page 388

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-237
F>A
Page:
O, S
Stand:
A
Abbreviation:
F
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. F>A ..... addition FEED -> AUX V.add ........mL
Function
Transfer solution from sample vessel (FEED) to the auxiliary vessel
(MEAS) on the 695 Autosampler.
Parameter
Range; Default value / Meaning
V.add
0.010...50.000 mL; 1.000 mL
Transfer volume
Solution volume that should be transferred from the sample vessel on
the sample rack of the 695 Autosampler to the auxiliary vessel.
Description
The
F>A
instruction transfers the solution volume specified under
V.add
from the sample vessel (FEED) to the auxiliary vessel (AUX) on the 695
Autosampler.
Comments
•
The
F>A
instruction can be used only in operation with the 695
Autosampler.
•
The auxiliary vessel (AUX) must be located on the sample rack
immediately after the sample vessel (FEED).
•
If the solution volume in the sample vessel <
V.add
, an error
message appears.
see also
A>M
,
F>M