7.4 Troubleshooting

746 VA Trace Analyzer / 747 VA Stand

7-37

No addition

•

Has the correct standard solution been used or is the concentration of the

solution too low: increase the volume of the standard addition or use a higher

concentration or reduce the sample amount accordingly.

•

If organic components are present: carry out a UV digestion or similar.

•

Could it be possible that the peak is not the peak which is being looked for?

•

Concentration of the analyte is too high: dilute.

7.4.3

Problems with HMDE

Low background current or unstable baseline

•

The electrode drops or the drops do not remain hanging: check MME.

Change capillary if necessary (see section 3.4.8) or replace sealing needle

(see section 3.4.9).

•

Has the sample been properly degassed?

•

Check the concentration of the electrolyte and the pH of the solution.

•

Check the initial potential and final potential of the analysis.

•

If the ion concentration in the solution is too high: dilute the electrolyte.

•

If the concentration to be determined is considerably higher than expected:

reduce sample volume or change the mode (e.g. from HMDE to SMDE or

DME).

•

Electrolyte solution too old: make up a new one. Its working life with organic

additives may be as short as 1 day or less.

Curves look like a "starry night"

•

Check MME. Check for break in contact between needle and capillary. If nec-

essary, clean the MME (see section 3.4.9).

•

If the electrode surface is overcharged: check deposition potential and time.

•

No drops at the capillary: change the capillary (see section 3.4.8) or replace

sealing needle (see section 3.4.9).

•

Have stirring and degassing been switched off during the measurement?

•

Check tapping mechanism on stand. If tapping strength is too weak, turn

slotted screw 71

71 on the valve block during operation in an anticclockwise di-

rection until a drop falls each time the tapper is triggered.

•

Electrolyte solution too old: make up a new one. Its working life with organic

additives may be as short as 1 day or less.