3.9 Connection of 685 Dosimats

746 VA Trace Analyzer / 747 VA Stand

3-33

3.9

Connection of 685 Dosimats

Up to four 685 Dosimats can be attached to the 746 VA Trace Analyzer for the

automatic addition of standard and auxiliary solutions. The 685 Dosimat and the

accessories needed have the following ordering designations (see also section

9.5):

• 2.685.0010

Dosimat

• 6.2134.000

Cable 685 – 746

• 6.3014.XXX Exchange unit, with PCTFE/PTFE flat stopcock

6.3014.153 burette volume V = 5 mL
6.3014.213 burette volume V = 10 mL
6.3014.223 burette volume V = 20 mL
6.3014.253 burette volume V = 50 mL

The choice of Exchange unit depends on the volume of liquid the Dosimat should

dispense. A burette volume of 5 mL is recommended for additions in the

µ

L range

(standard additions solutions), a burette volume of 10 mL or higher is recom-

mended for additions in the mL range (auxiliary solutions).
This section describes the connection procedure; further details on the 685 Dosi-

mat and the various Exchange units can be found in the 685 Instructions for Use of

the 685 Dosimat.

3.9.1

Electrical connection

Before a Dosimat is attached to the 746 VA Trace Analyzer, the 746 VA

Trace Analyzer must be switched off using the mains switch 11

11.

Connecting a 685 Dosimat to the 746 VA Trace Analyzer when the

latter is switched on can lead to destruction of the Dosimat!

The 685 Dosimat is connected to the 746 VA Trace Analyzer with the optionally

available 6.2134.000 Cable. The "remote control" sockets of the 685 Dosimat are

connected to the appropriate socket "Dos 1...4" (18

18 ... 21

21) of the 746 VA Trace

Analyzer (see Fig. 19). Maximum 4 Dosimats can be attached in this manner.
For settings (dosing rate, filling rate, etc.) and manual operation (manual filling or

dispensing) of the 685 Dosimat, dialog page "DOSIMATS" must be selected (see

section 5.3.3).

685

remote
control

0

I

Control Lines

Printer

External Bus

Dos. 1

Dos. 2

Dos. 3

Dos. 4

RS 232 Interface 1

RS 232 Interface 2

Reset

2A(TH)

115V

230V

230V

746

Fig. 19: Connection of 685 Dosimats to 746 VA Trace Analyzer

6.2134.000 cable