5.5.3 CURVES

746 VA Trace Analyzer / 747 VA Stand

5-111

Softkeys

Meaning

Scale

Axis scaling

Possibility for interactive modification of the parameters for the axis
scaling of the segment curves (display/plot parameters of the seg-
ments, see also section 5.6.2).

Unlinked

peaks

Non-assigned peaks

Display of all peak and wave voltages

U.ver/mV

found for the se-

lected segment but not linked to a substance.

Spike

elimin.

Spike elimination

Possibility for interactive modification of spike measured values.

This softkey is not available for curves recorded with

PSAMODE.

Parameter

Range; Default value / Meaning

Segment-VR-
Curve

display only

Identification of selected curve

Segment

Name of selected segment

VR

Variation and replication number of the selected
segment

Curve

original

Original curve

processed

Processed curve

smoothed

Smoothed and evaluated curve with baseline

derived

Derived curve (1

st

derivative)

-backgnd

Background curve (only with determinations with
BACKGND instructions, possible for all four curve
types)