219 (csweep, M, a – Metrohm 746 VA Trace Analyzer User Manual

Page 370

Advertising
background image

5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-219

(CSWEEP

Page:

S

Stand:

M, A

Abbreviation:

(CS

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. (CSWEEP

Function

Start of the block for a cyclic voltage sweep.

Description

The

(CSWEEP

instruction opens a block in a segment in which the ramps

for the cyclic voltage sweep are defined. This block can contain
maximum 5

RAMP

instructions, other instructions are not allowed. The

block must be closed with the

CSWEEP)

instruction.

Comments

•

Only one CSWEEP block is allowed per segment.

•

CSWEEP blocks may not be located within REP loops.

•

The electrode type (admissible are

HMDE

or

RDE

) and the measure-

ment mode

CYCMODE

must be entered before the CSWEEP block.

•

The number of measurement cycles for the CSWEEP block (max. 9)
is defined in the

CYCMODE

instruction by the

Meas.cycles

parameter.

•

Within a CSWEEP block a wait time between two ramps can be
programmed only by insertion of an additional, slow ramp as flat as
possible.

Example

see

RAMP

see also

CSWEEP), CYCMODE, RAMP