5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

985

Max. waiting time

If the signal drift has not yet been reached, then the measured value will
be accepted when the maximum waiting time has elapsed. If the waiting
time has not been newly entered then a waiting time that is suitable for
the drift will be calculated automatically according to the following equa-
tion:

Titrando, 855, 867
Input range

0 - 999999 s

Default value

52 s

Titrino
Input range

0 - 9999 s

Default value

52 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Titrando, 855, 867
Input range

0.1 - 999999.0 s (Increment: 0.1)

Default value

2.0 s

Titrino
Input range

0.08 - 16200 s (Increment: 0.08)

Default value

2.0 s

Stop measured value

Stop when the entered measured value has been reached since the start
of the measurement.

Titrando, 855, 867
Input range

-200.0 - 200.0 µA

Selection

off

Default value

off

off
No stop.

Measurement without drift control
If the option Measurement without drift control has been selected,
then the following parameters will be displayed:

Measuring time

Maximum period of time for measurement.