Metrohm tiamo 2.4 (process analysis) User Manual

Page 1469

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1453

Calibration range

Minimum and maximum concentration of the calibration solutions. In this
range, the calibration curve depicts a monotonously ascending or
descending function. The value cannot be modified. It is defined from the
method and corresponds to the range between the first and last calibra-
tion points used.

Calibration temperature

Temperature at which the calibration is carried out. In addition, the display
shows whether the temperature was measured with a Pt1000 or an NTC
temperature sensor or entered manually.

Input range

-20.0 - 150.0 °C

Default value

25.0 °C

NOTE

If the Calibration temperature has been entered manually, then the
following parameters will be reset to the default value or updated,
respectively:

Calibration temperature = manual

Calibration date is updated.

Calibration method = manual

Determination ID is deleted

User is updated

Audit Trail entry ensues

Calibration date

Display of date and time of the last calibration entered after each auto-
matic calibration or manual entry.

Calibration method

Display of the method name with which the last calibration has been car-
ried out. If the calibration data has been entered manually, then manual
will appear.

Determination ID

Identification of the determination.

The determination ID can be copied and pasted into another field.

User

Display of the short name of the user logged in during the calibration or
who entered the calibration data manually. If work is not carried out using

Advertising