5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

564

■■■■■■■■

tiamo 2.4 (for Process Analysis)

descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.

5.6.3.2.5.10

DET Upol - Volume window

Dialog window: Method

DET Upol Properties... Potentiometric evalua-

tion

Evaluation with volume window (mL) [New/[Properties] Volume win-

dow #

Volume windows are defined regions (windows) on the volume axis for
which different parameters for the potentiometric evaluation can be
defined. Only those endpoints which lie within these windows and addi-
tionally meet the parameters defined for each window will be recognized.
Only one endpoint will be recognized per window.

Lower limit

Lower limit for the volume window.

Titrando, 855
Input range

0.00000 - 9999.99 mL

Default value

0.00000 mL

Upper limit

Upper limit for the volume window.

Titrando, 855
Input range

0.00000 - 9999.99 mL

Default value

9999.99 mL

EP criterion

Criterion for the recognition of endpoints. Endpoints whose discovered
ERC value is smaller than the value entered here will not be recognized.

Titrando, 855
Input range

0 - 200

Default value

5

EP recognition

Filter for the recognition of equivalence points:

Titrando, 855
Selection

first | greatest | last | ascending | descending

Default value

first

first
Only the first endpoint to be found will be recognized.