Metrohm tiamo 2.4 (process analysis) User Manual

Page 33

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

1 Introduction

tiamo 2.4 (for Process Analysis)

■■■■■■■■

17

In the online help for Export configuration data and Import con-
figuration data
, the descriptions for Export templates and Elec-
trode type templates
were not available (the error was already elimi-
nated in tiamo 2.3, Patch 1).

If you cannot log in within the number of attempts specified in the
security settings and if at the same time the Message by e-mail
option is enabled, then an e-mail will be sent to the specified address.
Client name was output as text in the subject line of this e-mail, and
the associated value corresponded to the name of the computer;
Client name was replaced by Computer name (the error was
already eliminated in tiamo 2.3, Patch 1).

If the settings in the user administration were output as a PDF, then it
could happen that the PDF file did not correspond with the settings in
the software for some elements.

The input range for the titer value of a solution was modified in the
online help.

If the table of colorimetric sensors was sorted according to the date,
tiamo crashed.

At least one titrant with invalid titer existed in the Titrants/Solutions
subwindow. If the table was then sorted according to titer, then noth-
ing was displayed in this program part (empty display).

If a pH calibration or an electrode test was carried out with an intelli-
gent sensor on one system (tiamo and Titrando) and the sensor was
then transferred to another system, then the Calibration tempera-
ture
field on the Calibration data tab was empty.

If Monitor TC data was enabled for a sample solution (TC conductiv-
ity) and the validity for the TC data had expired, then no message was
displayed during the method run of a measurement.

Manual program part

If the measurement of a spectrum with the Avantes spectrometer was
canceled by clicking on [Stop], the [Stop] button correctly changed to
[Start], but the recording of the spectrum continued (the error was
already eliminated in tiamo 2.3, Patch 1).

When trying to approach a lift height greater than 90 mm with the
774 Oven Sample Processor, the program was unexpectedly finished.

A table with the gas flow factors for various carrier gases enabling the
correct determination of the gas flow in the 874 USB Oven Sample
Processor has been added to the online help.