6 sample changer - heater/gas, Heater/gas – Metrohm tiamo 2.4 (process analysis) User Manual

Page 1703

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

8 Manual control

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1687

Selection

On/Off | Duration

Default value

On/Off

Duration

Entry of the duration during which the pump/valve should remain
switched on. This field is only displayed if Operation = Duration.

Input range

1 - 999,999 s

Default value

60 s

[Start]

Starts pump 2. The time elapsed since the start is shown in the status dis-
play.

[Stop]

Stops pump 2.

8.4.4.6

Sample changer - Heater/Gas

Program part: Manual

The heater and gas flow on the 774 and 874 Oven Sample Processor can
be switched on and off here.

Heater
Heating the oven up to the desired temperature during the defined heat-
ing time.

Temperature

Temperature to which the oven has to be heated. With Init it is heated up
to the initial temperature set on the instrument.

Input range

50 - 250 °C

Default value

50 °C

Selection

Init

Heating period

Duration of the heating period until the desired temperature is reached.

Input range

1 - 999 min

Selection

off

Default value

off

[Start]

Starts the heater. The current temperature is shown in the status display.

[Stop]

Stops the heater.