Page 913
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4 (for Process Analysis)
■■■■■■■■
897
Titrando, 855
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
Sensor
Measuring input
Selection of the measuring input to which the sensor is connected.
Titrando
Selection
1 | 2
Default value
1
855
Selection
1
Default value
1
Titrino
Selection
1 | 2 | diff.
Default value
1
Sensor
Selection of a sensor of type pH electrode, metal electrode, pH elec-
trode or ISE electrode from the list of sensors defined in the sensor
table. The calibration data for the sensor will be adopted for pH and ISE
electrodes.
Selection
Sensor name | pH electrode | Metal electrode |
ISE electrode
Default value
Metal electrode
Temperature measurement
Type of temperature measurement.
Titrando, 855
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise the temperature entered manually
under Temperature on the Titration parameters tab will used.