5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1054

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Opens the Fixed endpoint evaluation # dialog window for entering a
new fixed endpoint (see Chapter 5.6.4.14.6, page 1057).

[Properties]

Opens the Fixed endpoint evaluation # dialog window to edit the
fixed endpoint selected in the table (see Chapter 5.6.4.14.6, page 1057).

[Delete]

Deletes the selected line.

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is activated, then the associated time for the minimum mea-
sured value will be interpolated from the measuring point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Input range

0.1 - 1,000.0 mAU/s

Default value

20.0 mAU/s

Maximum evaluation

Maximum evaluation

on | off (Default value: off)

If this option is activated, then the associated time for the maximum mea-
sured value will be interpolated from the measuring point list.

Threshold value

The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.

Input range

0.1 - 1,000.0 mAU/s

Default value

20.0 mAU/s