5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1217

not defined
In this case, the device name must be assigned by the user at the start
of the method.

Device type

Display or selection of the device type. If a device is selected under Device
name
, then this Device type field can no longer be edited, and the
device type belonging to the selected device is displayed instead

If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.

Selection

Device types | 814 USB Sample Processor

Default value

814 USB Sample Processor

Target

Tower

Selection of the tower of the Sample Processor for moving to the required
lift position. Both Tower 1 and Tower 2 can always be selected even if the
Sample Processor has only one tower.

Selection

1 | 2

Default value

1

Move

Selecting the target position on the rack:

Selection

Sample position | Rack position | Special
beaker
| Relative angle

Default value

Sample position

Sample position
Position defined for the current determination in the Run subwindow
of the program part Workplace as the Sample position parameter.
Rack position
Absolute position on the rack.
Special beaker
Special position on the rack defined in the rack properties of the Con-
figuration
program part.
Relative angle
The sample rack can be rotated by a certain angle relative to the cur-
rent position independently of rack positions. This can be used for
example to remove vial caps automatically.