Metrohm tiamo 2.4 (process analysis) User Manual

Page 1422

Advertising
background image

6.6 Subwindow Titrants/Solutions

■■■■■■■■■■■■■■■■■■■■■■

1406

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Next titer determination

Date on which the next titer determination is to be carried out. The date

can be selected by clicking on

in the Select date dialog window (see

Chapter 2.5.1, page 91). After the date has been entered, the Validity
field will be adapted automatically.

Default value

   Date titer det.+ 999 days

Message
The message options can only be edited if monitoring is switched on.

Message by e-mail

on | off (Default value: off)

If this option is enabled, the message is sent to the address defined under
[E-mail]. The message is sent in text format.

[E-mail...]

[E-mail...] opens the Send e-mail window (see Chapter 2.6.1, page 93).

Acoustic signal

on | off (Default value: off)

If this option is enabled, an acoustic signal will be emitted in addition to
the message defined above.

Action
The actions can only be edited if monitoring is switched on.

If during titer monitoring it is found that the validity period of the titer has
expired, then one of the following actions will be triggered automatically
during the start test:

Selection

Record message | Display message | Cancel
determination

Default value

Display message

Record message
The message that the validity period has expired will be automatically
saved with the determination.
Display message
A message is displayed and you can select whether you want to con-
tinue the run or cancel it. If the run is continued, then the message
that the validity period has expired will be automatically saved with the
determination.
Cancel determination
The ongoing determination will be automatically canceled. The follow-
ing message must be confirmed with [OK].