Metrohm tiamo 2.4 (process analysis) User Manual

Page 1666

Advertising
background image

7.20 IO controller

■■■■■■■■■■■■■■■■■■■■■■

1650

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Input parameters

Tolerance range

for control task = digestion unit
Input range

0 - Smallest distance between setpoint value
and 0% mapping or 100% mapping 'Map-
ping unit of the input or of the calculated
value'

Default value

1.0 'Mapping unit of the input or of the cal-
culated value'

for control task = free controller
Input range

0 - Smallest distance between setpoint value
and 0% mapping or 100% mapping 'Map-
ping unit of the input or of the calculated
value'

Default value

1 'Mapping unit of the input or of the calcu-
lated value'

Working range

for control task = digestion unit
Input range

0 - Smallest distance between setpoint value
and 0% mapping or 100% mapping 'Map-
ping unit of the input or of the calculated
value'

Default value

40.0 'Mapping unit of the input or of the cal-
culated value'

for control task = free controller
Input range

0 - Smallest distance between setpoint value
and 0% mapping or 100% mapping 'Map-
ping unit of the input or of the calculated
value'

Default value

10 'Mapping unit of the input or of the calcu-
lated value'

Output parameters

PMW period duration

for control task = digestion unit
Input range

0 - 9,999.0 s

Default value

1.0 s