1 Introduction

tiamo 2.4 (for Process Analysis)

■■■■■■■■

15

If a determination was carried out with the 089 Photometer, then the
blank value was not subtracted in the database from the raw data IME
and EME (initial and end measured values, respectively) associated
with the commands MEAS Opt and MEAS Opt Conc (this error was
already eliminated in tiamo 2.3, Patch 1).

If statistically linked determinations that also contained as results such
things as the slope, the axis intercept and the correlation coefficient for
linear regressions (variables SLO, ITC, COR) were recalculated in such
a way that a determination or a result was switched off, then the recal-
culated results of slope, axis intercept and correlation were not correct
(this error was already eliminated in tiamo 2.3, Patch 1).

If report templates with modified curve properties with regard to the
default templates supplied were imported from tiamo 1.x, then the set-
tings for the curve properties were reset to the default values after the
import.

The program occasionally crashed when generating a PDF with a
report template.

In tiamo 2.3, Patch 1, the information on the Thermoprobe was no
longer displayed on the Configuration tab in the Information sub-
window in the case of determinations involving thermometric titration.

If the ENTER key was pressed directly after the reprocessing dialog was
opened (without having made any modification), tiamo crashed.

The value of the ETR variables was not output correctly in the calcula-
tions.

In the commands STDADD dos and STDADD auto, no entries were
present for the parameters of the solution (solution name, concentra-
tion, production date, titer, date of titer determination and titer
method) on the Configuration tab in the Information subwindow.
In addition, the titer was not correctly stored.

Recalculating a thermometric determination in which the TET com-
mand was not run through (no measuring point list) caused a crash.

Method program part

An 867 pH Module present in the device configuration could in fact be
selected as Device type in the ELT MEAS command, but not as
Device name. This led to a situation where the electrode test could
not be carried out with the 867 pH Module (the error was already
eliminated in tiamo 2.3, Patch 1).

The device 846 Dosing Interface could not be selected in the PORT
command. This led to a situation where the PORT command could
not be performed with this kind of device, even though the firmware
supported this (the error was already eliminated in tiamo 2.3, Patch 1).

The description of Fill automatically at end was not correct in the
online help for the ADD command (the error was already eliminated in
tiamo 2.3, Patch 1).