Metrohm tiamo 2.4 (process analysis) User Manual
Page 1729
■■■■■■■■■■■■■■■■■■■■■■
8 Manual control
tiamo 2.4 (for Process Analysis)
■■■■■■■■
1713
Current port
Shows the current port position.
Port
Selection of the target port to be switched to.
Input range
1 - 12
Default value
1
Shift direction
Selection of the direction in which the valve moves to the target port.
Selection
ascending | descending | automatic | not over
Default value
ascending
ascending
The valve moves towards the target port in ascending direction.
descending
The valve moves towards the target port in descending direction.
automatic
The valve determines the shortest path to the target port on its own.
not over
The valve moves towards the target port in such a way that the port
specified in the parameter Not over will not be traversed.
Not over
Selection of that port which is not to be traversed when the port is
changed. This parameter can only be edited for Shift direction = not
over.
Input range
1 - 12
Default value
1
[Start]
Set the definite values.
Move to current port position +1.
Move to current port position –1.