5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1035

Switch off automatically

on | off (Default value: on)

If this option is activated, the stirrer will be switched off automatically
when the command has finished.

5.6.4.11.3

MEAS TC Cond - Measuring parameters

Tab: Method

▶ MEAS TC Cond ▶ Properties... ▶ Measuring parameters

Command name

Name of the command.

Entry

25 characters

The parameters for the measuring procedure can be defined on this tab.

Measurement frequency

712
Selection

auto | 300 Hz | 2.4 kHz

Default value

auto

auto
The more suitable of the two frequencies is selected.

Stop time

Maximum period of time for measurement.

Input range

0 - 999999 s

Default value

3000 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

712
Input range

0.4 - 999999 s

Default value

2.0 s

856
Input range

0.1 - 999999 s

Default value

2.0 s

Sample solution

Selection of the sample solution from the table Sample solutions (TC
conductivity)
.

Selection Table entries
Default value first entry of table