6.7 Sensors subwindow

■■■■■■■■■■■■■■■■■■■■■■

1430

■■■■■■■■

tiamo 2.4 (for Process Analysis)

Input range

–2,000.0 - 2,000.0 mV

Default value

0.0 mV

c (blank)

c (blank) is the third characteristic of the calibration function U = f(log
c)
. It reflects to a certain extent the curvature of the calibration function at
the lowest concentrations. This curvature is caused by the influence of so-
called interfering ions. If the concentration calibration is carried out with
less than 3 standards, then c (blank) will be set equal to zero. The c
(blank)
parameter is available for calculations as the variable 'Command
name.BLV'
.

Value
Input range

0.00 - 999,999,999

Default value

0.00

Unit
Entry

10 characters

Default value

'empty'

Selection

mol/L | mmol/L | ppm | % | g/L | mg/L | µg/L |
mg/mL | µg/mL

Specific calibration data for conductivity sensors

Cell constant

Cell constant of the conductivity sensor. This value can be determined
automatically with a calibration or entered manually. The cell constant is
available for calculations as the variable 'Command name.BLV'.

Input range

0.001 - 500 /cm

Default value

1 /cm

Specific calibration data for the Thermoprobe

a ... e

Display of the calibration coefficients ae.

Shared calibration data

Calibration temperature

Temperature during the calibration. In addition, the display shows
whether the temperature was measured with a Pt1000 or an NTC temper-
ature sensor or entered manually (is not being displayed for the Thermop-
robe).

Input range

–20.0 - 150.0 °C

Default value

25.0 °C