5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

861

Troubleshooting

Problem

Remedy

The control point is not main-
tained well. The measured value is
at times too high and at times too
low. The controller "oscillates".

■

Reduce the Max. rate and
possibly also the Min. rate.

■

Set the Dynamics higher.

■

Check if stirring is effective.

■

Arrange the electrode and
buret tip to an optimum.

■

Possibly use an exchange/dos-
ing unit with a smaller cylinder
volume (smaller volume incre-
ment per pulse).

The control point is not reached
for too long.

■

Set the Dynamics lower.

■

Increase the Min. rate.

■

Increase the Max. rate.

The control range is not reached
quickly enough.

Set the setpoint value with a SET
pretitration in such a way that the
control range is reached more
quickly.

Commands
Depending on the measured value one of the following two STAT com-
mands can be selected: