Client ID was output as text in various places (in the header of PDF
reports and the report templates, in the Information subwindow on
the Determination tab in the Database program part, in the Global
Variables
subwindow on the Global Variable tab in the Configura-
tion
program part). The associated value, however, corresponded to
the computer name. Client ID has now been replaced with Com-
puter name
(the error was already eliminated in tiamo 2.3, Patch 1).

Workplace program part

If a method was programmed in which a check at the time of the start
was required for one or several method variables of the type Number
(e.g. the Sample size) or Date/Time and which contained an ERROR
track, the exception wizard appeared in the Workplace program part
after the 014-152 Field empty message was exited when the deter-
mination was started (tiamo crashed) (this error was already eliminated
in tiamo 2.3, Patch 1).

If a method was written in which the file name did not have the exten-
sion .pdf in the REPORT command in the PDF file field (completely
forgotten or only the period forgotten, e.g. abc or abcpdf), then this
would lead to a crash at the start of a determination (the error was
already eliminated in tiamo 2.3, Patch 1).

If the digital output of the Avantes spectrometer was used (CTRL com-
mand) while a measuring command using the same device was also
present in the method, then the message 014-122 Device occupied
appeared (the error was already eliminated in tiamo 2.3, Patch 1).

In the View

Properties menu item, the properties of the respective

live display were always displayed as Properties Live display instead
of Properties Live display # (the error was already eliminated in
tiamo 2.3, Patch 1).

If it is determined during the start test that the wavelength of the cali-
bration of the colorimetric sensor does not match the wavelength of
the light source of the 089 Photometer in the MEAS Opt Conc com-
mand, then the message 014-042 Incorrect wavelength appears.
Instead of the wavelength value of the colorimetric sensor calibration,
the wavelength value of the light source of the 089 Photometer was
specified in the message (the error was already eliminated in tiamo 2.3,
Patch 1).

The stop conditions for the commands DIGITAL IN and ANALOG IN
did not function, although the maximum running time did (the error
was already eliminated in tiamo 2.3, Patch 1).

In the ELT MEAS command for the electrode test, the drift was not
evaluated at the correct time. This resulted in the sensor rating being
too strict.

If a measuring command contained in a separate track was called from
a loop, then the command's indexed variables were not available after
the first run.