Metrohm tiamo 2.4 (process analysis) User Manual
Page 537
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.4 (for Process Analysis)
■■■■■■■■
521
Input range
0 - 999999 s
Default value
1 s
Start volume
Start volume
Volume to be added before the start of the titration at the dosing rate
indicated.
Titrando, 855
Input range
0.00000 - 9999.99 mL
Default value
0.00000 mL
Titrino
Input range
0.00 - 999.99 mL
Default value
0.00 mL
Dosing rate
Speed at which the start volume is to be added. The maximum dosing rate
depends on the cylinder volume of the exchange unit or dosing unit used.
Titrando, 855
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Input range
0.01 - 150.00 mL/min
Selection
maximum
Default value
maximum
Start measured value
NOTE
Only displayed for Titrando and 855.
Start measured value
Once the start measured value is reached, the dosing of the start volume
is stopped and the next start condition is processed or the titration is
started. If the start measured value is achieved by the addition of a start
volume, then titration will start directly.
Input range
-2000.0 - 2000.0 mV