Metrohm tiamo 2.4 (process analysis) User Manual
Page 58
2.4 Formula editor
■■■■■■■■■■■■■■■■■■■■■■
42
■■■■■■■■
tiamo 2.4 (for Process Analysis)
Identifica-
tion
Description
Commands
.EP{x}.TIM
Time in s until the endpoint x (1 -
9) is reached
DET, MET, SET,
KFT, KFC, BRC
.EP{x}.VOL
Volume for the endpoint x (1 - 9)
in mL
DET, MET, SET,
KFT
.EPP
Position after finishing the com-
mand. 0 = invalid position.
PORT
.ETE
End temperature (temperature
following processing of the com-
mand) in °C
DET, MET, SET,
KFT, KFC, BRC,
STAT, MEAS
(without T/Flow,
Ref, Spec), CAL
Cond, CAL
MEAS, STDADD,
DOS, ELT MEAS
.ETR
Result of the electrode test.
Electrode quality:
0 = No data available
1 = Short circuit
2 = Excessive start drift
3 = Wrong buffer
4 = Problems with the diaphragm
5 = Reference system defective
6 = No buffer values for the tem-
perature or excessive deviation of
the temperature between buffers
7 = Partial short circuit
8 = Glass defective
9 = Poor electrode
10 = Usable electrode
11 = Good electrode
12 = Excellent electrode
ELT LOOP
.EVT
End volume (total dosed volume
at the end of the command) in
mL
DET, MET, SET,
KFT, STAT, DOS