5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1131

Moves the concentration downward (modifies sequence).

[New]

Add the concentration for the next standard.

[Properties]

Edit the concentration of the selected standard.

[Delete]

Deletes the selected concentration.

Request for standard exchange

on | off (Default value: on)

If this check box is activated, a request for manual solution exchange is
displayed after each time the calibration loop is run through. This parame-
ter must be disabled for automatic solution exchange when calibrating
with a sample changer.

Calibration curve

Calibration function

Selection of the function with which the calibration curve is to be calcu-
lated out of the calibration points.

Selection

linear | quadratic | cubic | automatic

Default value

linear

Confidence interval

Range within which the measured values for the determination of the cali-
bration curve must lie. Values outside this range are called outliers; they
are not integrated in the calculation of the calibration coefficient.

Selection

99% | 98% | 95% | off

Default value

98%

off
The confidence interval will not be checked.

5.6.5.11

CAL MEAS Opt

5.6.5.11.1

CAL MEAS Opt - Overview

Dialog window: Method

CAL MEAS Opt Properties... CAL MEAS Opt - 'Com-

mand name'

Command for measuring standard solutions for the calibration of
colorimetric sensors
, used together with the CAL LOOP Opt com-
mand.