4 meas u - evaluations, Evaluations – Metrohm tiamo 2.4 (process analysis) User Manual

Page 976

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

960

■■■■■■■■

tiamo 2.4 (for Process Analysis)

off
No stop.

Temperature

Temperature

Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously.

Titrando, 855, 867
Input range

-20.0 - 150 °C

Default value

25.0 °C

Titrino
Input range

-170.0 - 500 °C

Default value

25.0 °C

5.6.4.4.4

MEAS U - Evaluations

Tab: Method

MEAS U Properties... Evaluations

Command name

Name of the command.

Entry

25 characters

The following additional methods for evaluation of measurement curves
can be activated and defined on this tab.

Fixed endpoint evaluation

Fixed endpoint evaluation

on | off (Default value: off)

If this option is activated, then the associated values for the other mea-
sured quantities will be interpolated with a fix measured quantity (Mea-
sured value
or Time) for the fixed endpoint from the measuring point
list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).