5 Method

tiamo 2.4 (for Process Analysis)

■■■■■■■■

1167

Specification of volume and dosing time
You define the volume and the time in which this volume is to be dosed
(particularly suitable for synthesis applications). The dosing rate is auto-
matically optimized.

The time defined as Dosing time corresponds to the net dosing time t

1

+t

2

,

i.e. the filling time of the exchange or dosing unit is included in the total.

If the specifications with the maximum dosing rate cannot be achieved,
then dosing is carried out at the maximum rate. The following formula can
be used to estimate whether the maximum dosing rate is used or not:

Dosing rate = desired flow rate in mL/min • 2.22